Source.VolumeFlowRate.Gas

Source

This process sources a reference volume flow rate of gas for calibrating gas-flow meters. The instantaneous volumetric flow rate q_V equals dV/dt, sometimes estimated as ∆V/∆t using the total volume ∆V flowing through a defined surface in time ∆t.

Result

NameQuantitym-layer aspect
VolumeFlowRate——

Parameters

ParameterRequiredQuantitym-layer aspectDefinition or allowed values
VolumeFlowRateRequiredflow-volumevolumetric-flow-rate (AS79)Volume of substance passing through a specific cross-sectional area.
TemperatureRequiredtemperaturethermodynamic-temperature (AS101)Temperature of the gas used during the measurement.
PressureRequiredpressure:absolutepressure (AS14)Pressure of the gas used during the measurement.
RelativeHumidityRequiredrelative-humidityrelative-humidity (AS110)Relative humidity (water content) of the gas used during measurement, mostly applicable to ambient air as the medium.
CompressibilityRequired—null (AS1)Relative volume change as a result of a pressure change of the gas used.
ReferenceTemperatureRequiredtemperaturethermodynamic-temperature (AS101)Reference temperature condition to or from which to correct the flow rate.
ReferencePressureRequiredpressure:absolutepressure (AS14)Reference pressure condition to or from which to correct the flow rate.
ReferenceRelativeHumidityRequiredrelative-humidityrelative-humidity (AS110)Reference Relative Humidity: condition to or from which to correct the flow rate
ReferenceCompressibilityRequired—null (AS1)Relative volume change as a result of a pressure change for the reference gas.
AmbientTemperatureRequiredtemperaturethermodynamic-temperature (AS101)—
AmbientPressureRequiredpressure:absolutepressure (AS14)—
AmbientRelativeHumidityRequiredrelative-humidityrelative-humidity (AS110)—
OutletPressureRequiredpressure:absolutepressure (AS14)Outlet pressure ambient or otherwise.
ReynoldsNumberRequired—null (AS1)Dimensionless quantity that characterizes the degree of laminar versus turbulent flow.
GasVelocityRequiredspeedspeed (AS38)Speed of the gas through the measurement area.
GasSpeciesOptional——Species of source gas for flow calibrations Identifiers for elements, compounds, mixtures for flow calibrations. Air · He · Ar · N2 · CO2 · CH4 · C3H8 · C4H10 · H2S · CO · H2 · Cl2
InnerDiameterRequiredlengthlength (AS3)Inner diameter of the pipe used in the measurement.
DynamicViscosityRequireddynamic-viscositydynamic-viscosity (AS107)Fluid internal resistance to flow.