Source.VolumeFlowRate.Gas
Source
This process sources a reference volume flow rate of gas for calibrating gas-flow meters. The instantaneous volumetric flow rate q_V equals dV/dt, sometimes estimated as ∆V/∆t using the total volume ∆V flowing through a defined surface in time ∆t.
Result
| Name | Quantity | m-layer aspect |
|---|---|---|
| VolumeFlowRate | — | — |
Parameters
| Parameter | Required | Quantity | m-layer aspect | Definition or allowed values |
|---|---|---|---|---|
| VolumeFlowRate | Required | flow-volume | volumetric-flow-rate (AS79) | Volume of substance passing through a specific cross-sectional area. |
| Temperature | Required | temperature | thermodynamic-temperature (AS101) | Temperature of the gas used during the measurement. |
| Pressure | Required | pressure:absolute | pressure (AS14) | Pressure of the gas used during the measurement. |
| RelativeHumidity | Required | relative-humidity | relative-humidity (AS110) | Relative humidity (water content) of the gas used during measurement, mostly applicable to ambient air as the medium. |
| Compressibility | Required | — | null (AS1) | Relative volume change as a result of a pressure change of the gas used. |
| ReferenceTemperature | Required | temperature | thermodynamic-temperature (AS101) | Reference temperature condition to or from which to correct the flow rate. |
| ReferencePressure | Required | pressure:absolute | pressure (AS14) | Reference pressure condition to or from which to correct the flow rate. |
| ReferenceRelativeHumidity | Required | relative-humidity | relative-humidity (AS110) | Reference Relative Humidity: condition to or from which to correct the flow rate |
| ReferenceCompressibility | Required | — | null (AS1) | Relative volume change as a result of a pressure change for the reference gas. |
| AmbientTemperature | Required | temperature | thermodynamic-temperature (AS101) | — |
| AmbientPressure | Required | pressure:absolute | pressure (AS14) | — |
| AmbientRelativeHumidity | Required | relative-humidity | relative-humidity (AS110) | — |
| OutletPressure | Required | pressure:absolute | pressure (AS14) | Outlet pressure ambient or otherwise. |
| ReynoldsNumber | Required | — | null (AS1) | Dimensionless quantity that characterizes the degree of laminar versus turbulent flow. |
| GasVelocity | Required | speed | speed (AS38) | Speed of the gas through the measurement area. |
| GasSpecies | Optional | — | — | Species of source gas for flow calibrations Identifiers for elements, compounds, mixtures for flow calibrations. Air · He · Ar · N2 · CO2 · CH4 · C3H8 · C4H10 · H2S · CO · H2 · Cl2 |
| InnerDiameter | Required | length | length (AS3) | Inner diameter of the pipe used in the measurement. |
| DynamicViscosity | Required | dynamic-viscosity | dynamic-viscosity (AS107) | Fluid internal resistance to flow. |