Accuracy spec → tolerance limits
Turn a manufacturer accuracy spec into tolerance limits at a test point: % of reading, % of range, counts at a resolution, ppm, % of full scale, absolute terms and “whichever is greater” floors. It shows how it read each term, so you can check the interpretation, and converts the half-width to a standard uncertainty.
Accuracy spec · tolerance at a test point
v0.1.0 · GUM §4.3Tolerance limits4.9986 to 5.0014 Va = ± 0.0014 V · u = 0.000808 V (a/√3)
How the spec was read
- 0.02 % of reading (5 V) = 0.001 V
- 4 counts × 0.0001 V resolution = 0.0004 V
- total 0.0014 V
- Limits: 5 ± 0.0014 V = 4.9986 to 5.0014 V
- Standard uncertainty: spec read as a bound, rectangular, u = a/√3 = 0.0008082903769 V
a = sum of terms (or the largest, for “whichever is greater”) · limits = reading ± a · counts × resolution
Test vectors for this tool
The reference cases this calculator is tested against on every change. Use them to check your own software. All test vectors · Download JSON
| ID | Inputs | Expected | Tolerance | Source |
|---|---|---|---|---|
| SPEC-01 | reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | 0.0014 V | ± 1e-15 | Keysight Knowledge Center, "What is the relationship between the 'Count' and the accuracy of U3401A or U3402A Digital Multimeter?" (https://docs.keysight.com/kkbopen/what-is-the-relationship-between-the-count-and-the-accuracy-of-u3401a-or-u3402a-digital-multimeter-588280703.html): 0.02 % of reading + 4 counts at 5 V, 100 µV resolution = ±0.0014 V. Spec confirmed in Keysight data sheet 5990-3970EN (Aug 2017), U3401A 5 V DC range: 100 µV resolution, ±(0.02 % of reading + 4 counts), 1 year, 23 °C ± 5 °C. The article itself sits behind a bot check and has not been opened; confirm it in a browser. |
| SPEC-02 | reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = lower | 4.9986 V | ± 1e-12 | Keysight Knowledge Center, "What is the relationship between the 'Count' and the accuracy of U3401A or U3402A Digital Multimeter?" (https://docs.keysight.com/kkbopen/what-is-the-relationship-between-the-count-and-the-accuracy-of-u3401a-or-u3402a-digital-multimeter-588280703.html): 0.02 % of reading + 4 counts at 5 V, 100 µV resolution = ±0.0014 V. Spec confirmed in Keysight data sheet 5990-3970EN (Aug 2017), U3401A 5 V DC range: 100 µV resolution, ±(0.02 % of reading + 4 counts), 1 year, 23 °C ± 5 °C. The article itself sits behind a bot check and has not been opened; confirm it in a browser. |
| SPEC-03 | reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = upper | 5.0014 V | ± 1e-12 | Keysight Knowledge Center, "What is the relationship between the 'Count' and the accuracy of U3401A or U3402A Digital Multimeter?" (https://docs.keysight.com/kkbopen/what-is-the-relationship-between-the-count-and-the-accuracy-of-u3401a-or-u3402a-digital-multimeter-588280703.html): 0.02 % of reading + 4 counts at 5 V, 100 µV resolution = ±0.0014 V. Spec confirmed in Keysight data sheet 5990-3970EN (Aug 2017), U3401A 5 V DC range: 100 µV resolution, ±(0.02 % of reading + 4 counts), 1 year, 23 °C ± 5 °C. The article itself sits behind a bot check and has not been opened; confirm it in a browser. |
| SPEC-04 | reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = standardUncertainty | 0.0008082903768654761 V | ± 1e-15 | Hand derivation from the term definitions and JCGM 100:2008 §4.3.7 (bounds, rectangular) |
| SPEC-05 | reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"normal","k":2,"confidence":"95 %"}, unit = V, output = standardUncertainty | 0.0007 V | ± 1e-15 | Hand derivation from the term definitions and JCGM 100:2008 §4.3.3 (stated coverage factor) |
| SPEC-06 | reading = -5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = lower | -5.0014 V | ± 1e-12 | Hand derivation from the term definitions |
| SPEC-07 | reading = 40, resolution = 0.001, terms = [{"type":"percent","of":"reading","value":1},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = lower | 39.596 V | ± 1e-12 | Hand derivation from the term definitions |
| SPEC-08 | reading = 4, range = 6, resolution = 0.001, terms = [{"type":"percent","of":"reading","value":0.5},{"type":"counts","value":2}], distribution = {"kind":"rectangular"}, unit = V, output = upper | 4.022 V | ± 1e-12 | Hand derivation from the term definitions |
| SPEC-09 | reading = 1, range = 10, terms = [{"type":"percent","of":"reading","value":0.0035},{"type":"percent","of":"range","value":0.0005}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | 0.000085 V | ± 1e-18 | Hand derivation from the term definitions |
| SPEC-10 | reading = 1, range = 10, terms = [{"type":"ppm","of":"reading","value":20},{"type":"ppm","of":"range","value":5}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | 0.00007 V | ± 1e-18 | Hand derivation from the term definitions |
| SPEC-11 | reading = 100, fullScale = 300, terms = [{"type":"percent","of":"full_scale","value":0.25}], distribution = {"kind":"rectangular"}, unit = psi, output = halfWidth | 0.75 psi | ± 1e-14 | Hand derivation from the term definitions |
| SPEC-12 | reading = 100, terms = [{"type":"percent","of":"reading","value":0.05},{"type":"absolute","value":0.002}], distribution = {"kind":"rectangular"}, unit = mV, output = halfWidth | 0.052 mV | ± 1e-15 | Hand derivation from the term definitions |
| SPEC-13 | reading = 5, terms = [{"type":"percent","of":"reading","value":1}], orGreaterOf = [[{"type":"absolute","value":0.1}]], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | 0.1 V | ± 1e-15 | Hand derivation from the term definitions |
| SPEC-14 | reading = 20, terms = [{"type":"percent","of":"reading","value":1}], orGreaterOf = [[{"type":"absolute","value":0.1}]], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | 0.2 V | ± 1e-15 | Hand derivation from the term definitions |
| SPEC-15 | reading = 5, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | SpecInputError | exact | Metrology Bench rule: never guess a missing piece |
| SPEC-16 | reading = 5, terms = [{"type":"percent","value":0.02}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | SpecInputError | exact | Metrology Bench rule: never guess a missing piece |
| SPEC-17 | reading = 5000, range = 10, terms = [{"type":"percent","of":"reading","value":0.02}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth | OutOfRangeError | exact | Metrology Bench rule: refuse, don’t extrapolate |
| SPEC-18 | reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"normal"}, unit = V, output = standardUncertainty | SpecInputError | exact | Metrology Bench rule: never guess a missing piece |