Accuracy spec → tolerance limits

Turn a manufacturer accuracy spec into tolerance limits at a test point: % of reading, % of range, counts at a resolution, ppm, % of full scale, absolute terms and “whichever is greater” floors. It shows how it read each term, so you can check the interpretation, and converts the half-width to a standard uncertainty.

Accuracy spec · tolerance at a test point

v0.1.0 · GUM §4.3
Spec terms, all in the reading’s unit; ± values from the datasheet, blank if not used
Tolerance limits4.9986 to 5.0014 Va = ± 0.0014 V · u = 0.000808 V (a/√3)
How the spec was read
  • 0.02 % of reading (5 V) = 0.001 V
  • 4 counts × 0.0001 V resolution = 0.0004 V
  • total 0.0014 V
  • Limits: 5 ± 0.0014 V = 4.9986 to 5.0014 V
  • Standard uncertainty: spec read as a bound, rectangular, u = a/√3 = 0.0008082903769 V
a = sum of terms (or the largest, for “whichever is greater”) · limits = reading ± a · counts × resolution
Display rounding

Test vectors for this tool

The reference cases this calculator is tested against on every change. Use them to check your own software. All test vectors · Download JSON

IDInputsExpectedToleranceSource
SPEC-01reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth0.0014 V± 1e-15Keysight Knowledge Center, "What is the relationship between the 'Count' and the accuracy of U3401A or U3402A Digital Multimeter?" (https://docs.keysight.com/kkbopen/what-is-the-relationship-between-the-count-and-the-accuracy-of-u3401a-or-u3402a-digital-multimeter-588280703.html): 0.02 % of reading + 4 counts at 5 V, 100 µV resolution = ±0.0014 V. Spec confirmed in Keysight data sheet 5990-3970EN (Aug 2017), U3401A 5 V DC range: 100 µV resolution, ±(0.02 % of reading + 4 counts), 1 year, 23 °C ± 5 °C. The article itself sits behind a bot check and has not been opened; confirm it in a browser.
SPEC-02reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = lower4.9986 V± 1e-12Keysight Knowledge Center, "What is the relationship between the 'Count' and the accuracy of U3401A or U3402A Digital Multimeter?" (https://docs.keysight.com/kkbopen/what-is-the-relationship-between-the-count-and-the-accuracy-of-u3401a-or-u3402a-digital-multimeter-588280703.html): 0.02 % of reading + 4 counts at 5 V, 100 µV resolution = ±0.0014 V. Spec confirmed in Keysight data sheet 5990-3970EN (Aug 2017), U3401A 5 V DC range: 100 µV resolution, ±(0.02 % of reading + 4 counts), 1 year, 23 °C ± 5 °C. The article itself sits behind a bot check and has not been opened; confirm it in a browser.
SPEC-03reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = upper5.0014 V± 1e-12Keysight Knowledge Center, "What is the relationship between the 'Count' and the accuracy of U3401A or U3402A Digital Multimeter?" (https://docs.keysight.com/kkbopen/what-is-the-relationship-between-the-count-and-the-accuracy-of-u3401a-or-u3402a-digital-multimeter-588280703.html): 0.02 % of reading + 4 counts at 5 V, 100 µV resolution = ±0.0014 V. Spec confirmed in Keysight data sheet 5990-3970EN (Aug 2017), U3401A 5 V DC range: 100 µV resolution, ±(0.02 % of reading + 4 counts), 1 year, 23 °C ± 5 °C. The article itself sits behind a bot check and has not been opened; confirm it in a browser.
SPEC-04reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = standardUncertainty0.0008082903768654761 V± 1e-15Hand derivation from the term definitions and JCGM 100:2008 §4.3.7 (bounds, rectangular)
SPEC-05reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"normal","k":2,"confidence":"95 %"}, unit = V, output = standardUncertainty0.0007 V± 1e-15Hand derivation from the term definitions and JCGM 100:2008 §4.3.3 (stated coverage factor)
SPEC-06reading = -5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = lower-5.0014 V± 1e-12Hand derivation from the term definitions
SPEC-07reading = 40, resolution = 0.001, terms = [{"type":"percent","of":"reading","value":1},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = lower39.596 V± 1e-12Hand derivation from the term definitions
SPEC-08reading = 4, range = 6, resolution = 0.001, terms = [{"type":"percent","of":"reading","value":0.5},{"type":"counts","value":2}], distribution = {"kind":"rectangular"}, unit = V, output = upper4.022 V± 1e-12Hand derivation from the term definitions
SPEC-09reading = 1, range = 10, terms = [{"type":"percent","of":"reading","value":0.0035},{"type":"percent","of":"range","value":0.0005}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth0.000085 V± 1e-18Hand derivation from the term definitions
SPEC-10reading = 1, range = 10, terms = [{"type":"ppm","of":"reading","value":20},{"type":"ppm","of":"range","value":5}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth0.00007 V± 1e-18Hand derivation from the term definitions
SPEC-11reading = 100, fullScale = 300, terms = [{"type":"percent","of":"full_scale","value":0.25}], distribution = {"kind":"rectangular"}, unit = psi, output = halfWidth0.75 psi± 1e-14Hand derivation from the term definitions
SPEC-12reading = 100, terms = [{"type":"percent","of":"reading","value":0.05},{"type":"absolute","value":0.002}], distribution = {"kind":"rectangular"}, unit = mV, output = halfWidth0.052 mV± 1e-15Hand derivation from the term definitions
SPEC-13reading = 5, terms = [{"type":"percent","of":"reading","value":1}], orGreaterOf = [[{"type":"absolute","value":0.1}]], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth0.1 V± 1e-15Hand derivation from the term definitions
SPEC-14reading = 20, terms = [{"type":"percent","of":"reading","value":1}], orGreaterOf = [[{"type":"absolute","value":0.1}]], distribution = {"kind":"rectangular"}, unit = V, output = halfWidth0.2 V± 1e-15Hand derivation from the term definitions
SPEC-15reading = 5, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidthSpecInputError exactMetrology Bench rule: never guess a missing piece
SPEC-16reading = 5, terms = [{"type":"percent","value":0.02}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidthSpecInputError exactMetrology Bench rule: never guess a missing piece
SPEC-17reading = 5000, range = 10, terms = [{"type":"percent","of":"reading","value":0.02}], distribution = {"kind":"rectangular"}, unit = V, output = halfWidthOutOfRangeError exactMetrology Bench rule: refuse, don’t extrapolate
SPEC-18reading = 5, resolution = 0.0001, terms = [{"type":"percent","of":"reading","value":0.02},{"type":"counts","value":4}], distribution = {"kind":"normal"}, unit = V, output = standardUncertaintySpecInputError exactMetrology Bench rule: never guess a missing piece